| Model | 
                                8D52A、8D52B | 
                            
                            
                                | Mode | 
                                Monitoring Mode Sleep Analysis Mode | 
                            
                            
                                | Pulse Oximetry (SpO2)
                                 | 
                            
                            
                                | Display Range | 
                                0% to 100% (Resolution 1%) | 
                            
                            
                                | Accuracy | 
                                
                                    No motion: 70% to 100% : ±2%;  
                                    Low perfusion: 70% to 100% : ±3%
                                 | 
                            
                            
                                | Method  | 
                                Dual wavelength LED (Dual wavelength LED (660 nanometers @0.8mW and 905 nanometers
                                    @1.2mW; both as max average)) | 
                            
                            
                                | Pulse Rate (PR) | 
                            
                            
                                | Range | 
                                30 to 250bpm (Resolution 1bpm) | 
                            
                            
                                | Accuracy | 
                                
                                    No motion: ±1bpm or ±1%, whichever is greater  
                                    Low perfusion: ±3bpm or ±3%, whichever is greater
                                 | 
                            
                            
                                | Auxiliary Feature | 
                            
                            
                                | Perfusions Index (PI) | 
                                0.1~20%; resolution 0.1% | 
                            
                            
                                | SDNN | 
                                
                                    successive differences between normal pulse rate at unit of millennium  
                                    second for a data period 5 minutes (Calculated from P-P interval)
                                 | 
                            
                            
                                | R-MSSD | 
                                
                                    root mean square of successive differences between normal pulse rate at unit of
                                    millennium second for a data period 5 minutes (Calculated from P-P interval)
                                 | 
                            
                            
                                | Alarm System (not for Sleep Analysis
                                        Mode) | 
                            
                            
                                | Beat Sound | 
                                On/off. | 
                            
                            
                                | Alarm volume | 
                                Adjustable (1~3) | 
                            
                            
                                | Mute Alarm  | 
                                Use input key for temporary mute | 
                            
                            
                                | Alarm Silence Duration  | 
                                30, 90s ; Preset 90s | 
                            
                            
                                | Communication Interface | 
                            
                            
                                | Wireless  | 
                                Bluetooth Low Energy 4.0  | 
                            
                            
                                | Cable Interface  | 
                                USB 2.0 | 
                            
                            
                                | Battery | 
                            
                            
                                | Main power  | 
                                2x AAA Alkaline battery | 
                            
                            
                                | Wrist Band | 
                            
                            
                                | Size | 
                                Adjustable for wrist perimeter 15 ~ 21 cm | 
                            
                            
                                | Material | 
                                Silicon rubber | 
                            
                            
                                | Environmental | 
                            
                            
                                | Operating | 
                                Temperature 1º~ 40º C; Humidity 10% ~ 95% non-condensing | 
                            
                            
                                | Storage and Transport  | 
                                Temperature -25º~ 70º C; Humidity 10% ~ 95% non-condensing | 
                            
                            
                                | Atmospheric Pressure | 
                                700 hPa ~ 1013 hPa for both operating and storage  | 
                            
                            
                                | Ingress Protection | 
                                IP55 |